| Appendix
2. Additional Information About Proposed Approach |
2.1
Estimated Annual Average PM concentrations
Pittsburgh
is an area influenced by reasonably well-defined local, remote, and
biogenic sources (Figure A.1). This will make source-receptor relationships
much easier to quantify compared to areas closer to the eastern seaboard
where the urban centers interact with each other strongly. The estimated
annual average PM2.5 concentration for the Supersite location
is approximately 20 µg m-3 (Falke, 1999). The maximum daily average
measured PM10 concentration in the area in 1998 was approximately
150 µg m-3.

Figure
A.1 Annual average PM2.5 concentrations (1994-1996) derived from PM10
aided spatial estimation, Falke (1999).
2.1
Acronyms A list of definitions of acronyms for instrumentation (see
Table 2) follows:
1 Ultrafine
SMPS: Scanning Mobility Particle Spectrometer (TSI model 3936N25) for
0.003-0.150 mm
2 SMPS:
Scanning Mobility Particle Spectrometer (TSI model 3934L) for 0.01-1.0
mm
3 APS:
Aerodynamic Particle Sizer Spectrometer (TSI model 3320) for 0.5-1.0
mm
4 ELPI:
Electrical Low Pressure Impactor (Dekati)
5 Ultrafine
CPC: Ultrafine Condensation Particle Counter (TSI model 3025A) for 0.003-10
mm
6 FRM:
Federal Reference Method PM2.5 Sampler
7 TEOM:
Tapered Element Oscillating Microbalance PM2.5 Sampler
8 CMU
Sampler: Carnegie Mellon University system shown in Figure 2
9 LPI:
Low Pressure Impactor developed by Susanne Hering
10
MOUDI: Micro-Orifice Uniform Deposit Impactor
11
IC: Ion Chromatography
12
ICPMS: Inductively Coupled Plasma Mass Spectrometer
13
OC and EC: Organic Carbon and Elemental Carbon
14
PC-BOSS: Particle Concentrator-Brigham Young University Organic Sampling
System. Analysis is conducted for several species: sulfate and nitrate
(IC), NH4+ (spectrophotometry), OC/EC (Temperature Programmed Volatilization
analysis), pH and acidity.
15
R&P sampler: Ruprecht and Patashnik sampling system
16
GC-MS: Gas Chromatography-Mass Spectrometry
17
FTIR: Fourier Transform Infrared Spectroscopy
18
MOUDI-PUF system: MOUDI-Polyurethane Foam plug preceded by a cyclone
and denuder
19
HPLC: High Performance Liquid Chromatography
20
RSMS-II: University of Delaware Single Particle Mass Spectrometer
21
SEM: Scanning Electron Microscopy
22
HFAS/GFAA: High Frequency Aerosol Sampler/Graphite Furnace Atomic Absorption
Spectrophotometry
23
TDMA: Tandem Differential Mobility Analyzer
24
CCN Counter: Cloud Condensation Nuclei Counter manufactured by DH Associates
25
GC-FID: Gas Chromatography-Flame Ionization Detector
26
CSU collector: Colorado State University cloudwater collector
2.2
Speciation of Organic Aerosol
The
material collected in the denuder, the filter and the PUFs will be analyzed
by an improved GC/MS technique developed by the Rogge group for the
quantification of the concentrations of individual compounds. The proposed
improved technique (for details see section 4.3) will allow the increase
of the pool of organic compounds that can be chromatographically separated
and made elutable from the gas chromatographic column used. Daily filter
samples will be collected throughout the sampling campaign. Collected
filter samples will be grouped into biweekly composites and analyzed
together, yielding biweekly average ambient fine particle organic compounds
concentrations. Each 29th day, XAD-Denuder/Filter/PUF sampling will
be conducted and the entire set will be analyzed separately to obtain
additional detailed information about the partitioning of semi-volatile
organic compounds between the gas- and particle-phases. During the intensive
sampling periods the sampling frequency will be increased to two samples
per day (12 hour samples) and the XADs, filters, and PUFs each will
be analyzed separately. For 2 days of two intensive campaigns, 4 hours
sampling will be conducted and all XADs, filter, and PUFs analyzed separately.
The proposed schedule will provide a unique dataset with high temporal
and chemical resolution that will broaden the knowledge about the chemical
composition of organic particulate matter. This dataset will allow the
Rogge group to derive source/receptor relationships using their published
organic source profiles together with the collected ambient data.
Each
of the samples (XAD, filter, PUF), and composited filter samples will
be spiked with 7 deuterated recovery standards before extraction. XAD
will be extracted successively four times with a mixture of dichloromethane/acetone/hexane
(2:3:5) (Gundel et al., 1995). Quartz fiber filters will be extracted
with dichloromethane three times using mild sonication. Similarly, PUF
samples will be extracted three times with dichloromethane in a specially
designed apparatus that allows to repeatedly compress the PUF foam plug.
After extraction, the sample volume will be reduced to about 0.5 to
2.0 ml and divided into two aliquots. One aliquot will be used for derivatization
with diazomethane to convert organic acids to their methyl ester analogues
and acidic hydroxy compounds to their respective methoxy analogues.
After the derivatization, both sample aliquots will be stored in a freezer
until analysis by GC/MS. In order to make other polar organic compounds
accessible to chromatography, a method will be developed and refined
during this project that uses BSTFA [bis(trimethylsilyl)trifluoroacetamide]
for compound derivatization by silylation (Simoneit et al., 1999). Furthermore,
for selected samples, silica gel column chromatography will be conducted
with 8 different eluting solvents, fractionating the sample extract
according to compound polarity. Solvent elution sequence includes: (F1)
hexane, (F2) hexane/toluene (3:1), (F3) hexane/toluene (1:1), (F4) hexane/ethyl
acetate (19:1), (F5) hexane/ethyl acetate (9:1), (F6) hexane/ethyl acetate
(6:1), (F7) hexane/ethyl acetate (4:1), and (F8) methanol. In order
to obtained quantifiable fractions, a method will be developed that
allows to quantitatively track the fractionation and conversion of the
polar fractions into GC chromatographicable derivatives.
Extract
analysis will be accomplished using a Hewlett-Packard Model 5973 quadrupole
mass spectrometer coupled with a HP 6890 series gas chromatograph. At
least the following compound classes will be searched for and when identified
also quantified: n-alkanes, n-alkanoic acids, n-alkanols, n-alkanones,
furans, furanones, resin acids, levoglucosan, sterols, PAHs, compounds
identified in smog chamber experiments as possible secondary organic
atmospheric reaction products, and others. As far as available, high
purity standards will be used to aid quantification. Beyond the listed
compounds, every effort will be made to identify in a few selected samples
everything possible, applying the extract derivatization and fractionation
methods to be developed.
2.3
Size-resolved FTIR spectra with solvent rinses
Size-segregated
samples will be collected on five impaction stages with cut points between
0.05 and 2.0 mm for 24 hours on ZnSe disks at 1 lpm using a Hering Low
Pressure Impactor (LPI) during the three intensive periods. Spectra
will be taken of a clean area of the ZnSe disk and of the deposit using
a Mattson Research Series 100 FTIR spectrometer, and spectral results
will be presented as the percent of incident light absorbed versus wavenumber.
Each spectrum will be an average of 200 spectral scans at 2 cm-1 wavenumber
resolution. Functional groups will be identified based on spectral libraries
and standards. Instrumental sensitivity will be evaluated by daily analysis
of a standard thickness polystyrene film and weekly analysis of KBr
pellets comprised of mixtures of organic and ammonium sulfate standards.
Following initial analysis samples will be gently rinsed with spectral-grade
hexane, spectral-grade acetone, and deionized water and reanalyzed after
each rinse. The rinse is performed by holding the substrate at a 45°
angle and applying 1 ml of solvent to the deposit from a pipet. The
solvent is allowed to flow slowly over the deposit for approximately
5-10 seconds. This process will be repeated twice. The substrates are
then air dried and reanalyzed by FTIR. Acetone is applied to the sample
after the final water rinse to facilitate drying. Spectra are integrated
using GRAMS 32 peak integration software and functional group absorbances
are quantified relative to sulfate.
2.4
LIBS
Laser-Induced
Breakdown Spectroscopy (LIBS) is an optical technique that can be used
to detect trace inorganic species and particles at femtogram-picogram
levels in combustion products and in the ambient atmosphere. LIBS has
been used as an analytical technique for gases, liquids, and solids
(Song et al., 1997; Schechter, 1997). Applications of LIBS employ a
pulsed laser with a high peak power to form a spark (breakdown) in the
medium to be examined.
 |
| Figure A2:
Conceptual illustration of the LIBS system and a sample spectrum. |
In
gases, the temperature of the resulting plasma is in the range of 10,000
- 15,000 K, hot enough to fragment all molecules into their constituent
atoms, and to excite the electrons in the atoms out of the ground state
and into excited electronic states. As the plasma cools, excited electrons
relax back into their ground states, emitting light at characteristic
atomic frequencies. Identification of the atoms present in the sample
volume occurs using well-known atomic emission lines, and quantification
of the atomic species concentration occurs via quantification of the
intensity of the atomic emission lines. A conceptual diagram and sample
spectra are illustrated in Figure A2. Sensitivities range from femtograms
for some species, such as Be, to well above the picogram level for other
species, such as Sb.
LIBS
is extremely well suited as a diagnostic for trace inorganic species.
The Buckley group has recently been directing a project at Sandia National
Laboratories to apply LIBS as a real-time, in situ monitor for toxic
metals in combustion systems (Buckley et al., 1999; Hahn et al., 1999).
A field detection system has been assembled that can run in an unmanned
mode for continuous emissions monitoring (Figure A2). Laboratory experiments
suggest that LIBS can be used to measure particles as small as 100-150
nm depending on the species present. With the capability in the software
to detect single particle hits, it is possible to do joint composition
and sizing analysis of single particles. This ability makes LIBS an
excellent diagnostic for trace species in the ambient atmosphere. Laboratory
measurements to validate this technique have been accomplished. Ambient
measurements have been made at industrial sites and in Livermore, CA.
2.4
Single Particle Mass Spectrometry
There
are now a number of groups world-wide who have built single particle
instruments and employed them in field campaigns (e.g., Noble and Prather,
1996; Murphy et al., 1998). All of these measurements were performed
using the light scattering type of instrument so only examined particles
greater than a few hundred nanometers in diameter. From experience with
single particle instruments and a large body of literature on laser
ablation mass spectrometry, we can summarize the capabilities of the
current RSMS-II design that are especially suited for exploring atmospheric
aerosols.
1.
External mixing properties -- atmospheric particles are emitted
by a wide range of sources and undergo disparate processing in the atmosphere.
This results in particles of a given size having a range of compositions.
These compositions can often be grouped into categories or classes of
similar composition (e.g., Song et al., 1999).
2.
Composition of fine and ultrafine particles
-- Although there are large numbers of fine and ultrafine particles
in the atmosphere, the mass of particles in these size ranges is quite
small. As a result of this small mass loading, using conventional techniques
to analyze the particles in these size classes necessitates sampling
a large volume of air either at a high volume flow rate or over a long
sampling period. The single particle technique described here analyzes
individual particles so the overall particulate mass in a given size
is not a constraint. Wexler and Johnston have demonstrated that single
particle analysis is effective down to about 10 nm (Carson et al., 1997)
and has good sensitivity to impurities at 50 nm (Ge et al., 1998). This
particle size range is extremely important to atmospheric processes
such as cloud dynamics and radiative balance.
3.
Size distribution of composition
-- In all the experiments proposed here, RSMS-II will be run in parallel
with independent particle sizing instruments (a TSI SMPS and APS) so
that the particle hit rate can be normalized to the actual particle
concentration that is sampled. Thus particle-sizing instruments will
record the size distribution while RSMS-II will be used to establish
the composition of particles analyzed at a given size. The RSMS-II hit
rate will be calibrated beforehand using particles that are easy to
analyze. Thus the discrepancy in the particle number distribution obtained
from RSMS-II and the pure sizing instruments will indicate the percentage
of particles either not analyzed or not analyzable.
4.
Speciating inorganic compounds -- The speciation abilities
of the instrument are currently not as varied or quantitative as bulk
instruments, but Wexler and Johnston have shown that for instance different
sulfur-containing compounds, such as sulfate, methane sulfonates, and
hydroxymethane sulfonates, can be distinguished from each other (Neubauer
et al., 1996). Elemental and metals analysis has also been demonstrated
(Murphy et al., 1998; Ge et al., 1998).
5.
Distinguishing elemental from organic carbon -- From the
laser ablation literature, it is known that elemental carbon particles
can be distinguished from organic carbon particles by the pattern of
carbon-hydrogen peaks in the mass spectra (DeWaele et al., 1983). The
usual laser desorption/ionization single particle analysis is only effective
for speciating aromatics and distinguishing organic from elemental carbon.
More in-depth single particle organic analysis techniques are described
later in this proposal.
6.
Temporal resolution
-- Atmospheric particles often contain labile compounds whose concentration
is affected by reactions or condensation/evaporation. Due to the presence
of point sources, fronts, and turbulent eddies, the concentrations and
compositions of atmospheric particles may exhibit sharp spatial gradients.
Single particle analysis is fast compared to bulk analytical techniques.
The transit time from sampling to analysis is measured in seconds so
that a) rapid gradients in number or composition can be identified and
b) labile compounds are not significantly transformed between sampling
and analysis. The anticipated hit rate of hundreds of particles per
minute will enable rapid transients in composition to be resolved.
Focusing
of Aerosol Particles. It has been known for a number of years
that passing an aerosol through a sharp or conical orifice where the
flow is choked (that is, sonic) focuses certain particles (Dahneke,
1982; Fernandez de la Mora and Riesco-Chueca, 1988 and references therein).
The focussed particles have a Stokes number value around one, the exact
value depending on the nozzle geometry and the distance from the nozzle
to the focal point. The U. of Delaware team employs this principle in
their current single particle instrument, a second-generation version
called RSMS-II (Rapid Single-particle Mass Spectrometry-II). The instrument
focuses only a narrow range of particles sizes to the source region
of the mass spectrometer, but this size range can be selected by adjusting
the pressure upstream of the nozzle. If we consider a given nozzle geometry,
let us term the Stokes number that is focused by Stkf. Thus we can write
Stkf = (Dp2rpuo/18mDn)Cc where Dp is the particle diameter that is focussed,
rp is its density, uo is the velocity through the orifice (which is
sonic since the flow is choked), m is the viscosity of air, Dn is the
orifice diameter and Cc is the Cunningham non-slip correction factor.
All the terms except Dp and Cc can be combined into an effective diameter,
Dp,max = (18mDn Stkf/ rpuo)1/2 giving (Dp/Dp,max)2 = 1/Cc. Since the
maximum value of Cc is 1, Dp,max is the maximum particle aerodynamic
diameter than the nozzle can focus. Cc is a function of the particle
diameter and pressure. The usual function for the Cunningham correction
factor is difficult to manipulate analytically (Seinfeld and Pandis,
1998), but a simple linear approximation, Cc = 1 + 1.66 (2l/Dp), is
close to the more precise equation and has a maximum error of only 10%
at Kn=1. Using this approximation for Cc gives a quadratic in Dp whose
solution is
Dp
= ((3.32l)2 + Dp,max 2)1/2 - 3.32l
where
l, the mean free path, is a function of pressure via l = lopo/p, where
po and lo are the pressure and mean free path at standard conditions.
Since the mean free path is a function of pressure, the particle size
focussed in the mass spectrometer can selected by adjusting the pressure
as long as this size is smaller than Dp,max.
The
maximum size that can be focussed, Dp,max, is a function of properties
of air (m, uo) and the particle (rp) which are not adjustable. The only
adjustable parameter is the nozzle diameter, Dn, and geometric considerations
such as the cone angle and distance from the orifice to the focal point,
which are incorporated in Stkf.
It
is best to run the orifice choked because this focuses the smallest
particles, but this requirement places limits on the orifice diameter
in that the vacuum pump driving the orifice must be large enough to
choke the flow. A 600 lpm pump is large but reasonably transportable
and leads to a 3 mm orifice. The result is that for unit density particles,
Dp,max is about 2 micron. At low pressures, the focussed diameter is
a linear function of pressure but this sensitivity reduces as the focussed
diameter approaches Dp,max. By scanning the pressure from about 300
torr to about 1 torr, the nozzle is able to focus particles ranging
from 1 micron to 10 nm in aerodynamic diameter - selecting the pressure
is the next step.
One
challenge to the design of the instrument is controlling the pressure
upstream of the nozzle over a wide dynamic range while efficiently transmitting
particles. We use a bank of 10 critical orifices whose flow is controlled
by a rotary valve (Valco Instruments http://www.vici.com).
The area of each orifice is selected to evenly distribute the 10 particle
diameters logarithmically over the 10 nm to 2 micron size range. Each
flow-control orifice is a sharp hole minimizing particle deposition
and associated clogging. After the orifice bank is a section of straight
tube where the flow straightens and becomes laminar before it enters
the focussing orifice, giving more reliable focussing characteristics.
The transmission efficiency of the inlet and the instrument's ability
to size and analyze atmospheric particles has been submitted for publication
(Mallina et al., 1999).
Standard
software in the University of Delaware laboratory will be used to convert
the time-based raw spectra to mass-based spectra that can then be interpreted.
The pressure will be converted to particle aerodynamic diameter via
equation 2. The parameter Dp,max is a function of the geometry of the
nozzle, which is fixed during each experiment. The pressure governs
the mean free path, l, which then determines the particles size that
is focussed. The pressure diameter relationship will be calibrated with
standard aerosols selected by a DMA both before and after each sampling
period. When aerodynamic techniques are employed to size particles,
as RSMS-II does, the particle density and aerodynamic properties contribute
to the uncertainty in sizing. The density can be estimated from the
particle composition reducing this uncertainty. Likewise, from the relative
humidity and composition, the particle shape (e.g., spherical liquid,
facetted crystal) can be estimated and used to reduce the uncertainty
in converting aerodynamic to physical diameter.
Particle
Detection. Particles that are focussed to the source region
of the mass spectrometer are detected in one of two ways. Larger particles
are detected by light scattering. A CW doubled Nd-YAG laser passes through
the particle beam just upstream of the center of the source region.
Two PMTs in the source region detect forward scattered radiation at
angles of +/- 30 degrees and signal coincidence from these two PMTs
is used to trigger the ablation laser (Ohigashi et al., 1994). Tests
show that light scattering is effective down to about 200 nm.
For
smaller particles, light scattering is not effective so the ablation
laser is free fired (Reents et al., 1994; Carson et al., 1997; Ge et
al., 1998). Currently we use an MPB PSX-100 laser, which fires at up
to 100 Hz. To maintain single particle analysis, it is important to
operate the instrument such that most laser shots miss a particle. For
instance, if one in 10 shots result is a hit, then 1 out of 100 hits
is a double hit, that is, two particles were in the source region. Here
the nozzle flow rate characteristics become useful. Generally, the atmosphere
contains a great many ultrafine particles and relatively fewer fines.
This is somewhat counterbalanced by the flow rate through the nozzle.
Larger particles are focussed at higher pressures where the sampling
flow rate is higher whereas smaller particles are focussed at lower
pressures where the sampling flow rate is lower. Nevertheless, if the
hit rate is too high, dilution of particle stream is necessary to maintain
single particle analysis. Particles are "detected" by the presence of
a spectrum. The technique has been used in the laboratory to analyze
particles down to 10 nm (Carson et al., 1997) and we have used it to
demonstrate sensitivity to 0.1 mass percent impurities in 50 nm particles
(Ge et al., 1998).
Since
some particles may not be readily analyzed by RSMS-II and for the smallest
particles, the spectrum is used to identify the presence of the particle,
the measurements contain some uncertainty as to the number of particles
present. As a result, the particle number will be measured simultaneously
with an independent instrument (a TSI SMPS). The RSMS-II and standard
number concentration measurements will be compared to determine the
fraction of the particles analyzed.
Particle
Analysis. The source region of the mass spectrometer is specially
designed to maximize the probability of a particle hit when the laser
is free-fired. First, the excimer laser beam propagates collinear to
and in the opposite direction from the particle beam. In most designs
these two beams are normal to each other so the overlap region is only
a few hundred microns. With collinear beams, the overlap is very large
and the source region size is governed by the ion optics in the mass
spectrometer. In RSMS-II the ion optics have been designed to allow
a 4 cm source region (Carson et al., 1997). The excimer laser beam is
focussed to the center of the source region where it has a waist about
0.6 mm in diameter. This widens to about 2mm at the top and bottom of
the source region. The particle beam passes through a 1 mm skimmer before
it enters the source region of the mass spectrometer. In the center
of the source region, 18 cm from the primary orifice, the beam is very
well defined and about 4 mm across (measured by impacting oleic acid
particles on a glass slide). It is the overlap between the hour-glass-shaped
laser beam and the conical particle beam which determines the particle
hit-rate probability. The overall particle hit rate is the product of
this probability and the volume flow rate sampled.
Data
System. The data system consists of a 500 MHz 8-bit A/D converter
mounted in a PC (Precision Instruments 9847). The board was customized
by the manufacturer so that valid spectra could be detected in firmware
on the board -- PCs are not yet fast enough to check in software each
spectrum for validity at 100 Hz. Valid spectra are stored on the PC
disk and archived on writeable CD-ROM.
Individual
spectra can be viewed in real time in the field and can be used provide
feedback to other measurements. Ultimately, a more statistically significant
analysis of the data will be performed. Specifically, spectra from each
particle each size will be grouped into categories of similar particle
type. Two methods will be employed. We have used the CART (Classification
And Regression Tree) algorithm successfully to bin laboratory generated
particles by composition and quantify the amount of different sulfur
compounds in a particle (Neubauer et al., 1996). To quantify, the algorithm
must be trained with standards of known composition, which may be impractical
for atmospheric applications. Alternatively, a neural network based
algorithm has been tested on atmospheric particles and is able to classify
them without a priori training (Song et al., 1999). This is clearly
advantageous for atmospheric applications. Alternatives to these algorithms
will be explored if they fail to classify the spectra.
RSMS-II
has a number of advantages over the more conventional designs currently
in use.
1.
Particle size selection. Particles can be sized and selected
over a much wider size range and the particle transmission rate for
the selected size is much higher than for nozzles that transmit a broader
range.
2.
Particle sampling. Instruments that transmit a range of sizes
must cope with the dramatic increase in particle number as the particle
diameter decreases. Since the vast majority of atmospheric particles
are small, significant time will be spent waiting for the larger particles
to appear. With the new design, a given size is sampled until a statistically
significant number are obtained and then another size is selected.
3.
Particle detection. For smaller particles where the atmospheric
number concentrations are typically higher, free firing is practical
for analyzing particles. The particle number concentration is so low
for larger particles that the hit rate becomes prohibitively low so
the particles are detected by light scattering.
4.
Particle size limits. In free-firing mode, the instrument is
not limited by light scattering for particle detection. Thus ultrafine
particles can be sized and sampled.
5.
Particle analysis. Impactors can be used to obtain aerosol composition
but the mass of smaller particles is very low in the atmosphere. Thus
a large sampling time is needed to obtain the composition of fine and
ultrafine particles for typical atmospheric loadings. Single particle
analysis is well suited to analyzing small particles because the analysis
is proportional to number not mass.
6.
Volume flow rate. The volume flow rate into the instrument is
proportional to the pressure upstream of the focussing orifice. As a
result this flow rate is higher for larger particles (which are fewer)
and lower for smaller particles (which are more plentiful) helping to
equilibrate the hit rate over particle size.
2.5
Cloud and fog water collection
The
Colorado State University's CASCC2 collects fog/cloud drops by drawing
droplet-laden air at 5.8 m3 min-1 over six banks of 508 µm diameter
Teflon strands. Drops are collected by inertial impaction on the strands.
The 50% size cut for the CASCC2 corresponds to a drop diameter of 3.5
µm. Collected drops coalesce and are carried by gravity and aerodynamic
drag down into a Teflon sample trough and through a Teflon sample tube
to a polyethylene sample bottle. The collector is cleaned prior to each
fog/cloud event and blanks are taken. Operation of the CASCC2 will be
automated for this study. Fog/cloud presence will be detected using
a Gerber Scientific Model PVM-100 Particulate Volume Monitor. This instrument
monitors forward light scattering to determine cloud/fog liquid water
content (LWC). When the LWC is sufficiently high to indicate fog presence,
the CASCC2 will automatically be turned on. At the same time the PVM-100
will activate a modem to place a call to a pager carried by the individual
conducting the fog measurements, notifying her/him that sampling has
commenced. The operator will come to the site immediately to ensure
equipment is operating properly and to change samples as needed.
2.6
Determination of semi-volatile material
The
combination of the technology used in the BIG BOSS sampling system (Tang,
1994) and the Harvard particle concentrator (Sioutas, 1994a,b) has resulted
in the Particle Concentrator-Brigham Young University Organic Sample
System (PC-BOSS) (Ding 1997, 1999a; Eatough 1999) (Figure A3). The PC-BOSS
has five advantages over diffusion denuder systems previously used to
determine fine particulate semi-volatile material.
1.
Both semi-volatile ammonium nitrate and semi-volatile organic compounds
are determined with the same sampler (Ding 1999b, Eatough 1999).
2.
The use of a particle concentrator eliminates the need for the independent
measurement of the efficiency of collection of gas phase organic compounds
by the diffusion denuder (Ding 1999a; Eatough 1999).
3.
The elimination of the independent measurement of denuder breakthrough
(Cui 1998; Eatough 1995; Tang 1994) results in a sampler which can be
used in routine field operation.
4.
The collection of the concentrated particles in a low-volume flow stream
simplifies the power requirements of the sampler.
5.
The effective high sample flow obtained with the particle concentrator
allows for the collection of samples over time periods as short as one
hour (Eatough 1999; Pang 1999a,b).
The
PC-BOSS sampler has been validated in sampling programs completed in
Tennessee (July 1997), in Riverside CA (August-September 1997), in Bakersfield
CA (February-March 1998) and in Provo UT (November-December, 1998) (Pang
1999a; Ding 1999b; Pang 1999b). In the experiments completed to date
the efficiency of the denuder for the removal of gas phase organic material,
nitric acid and SO2 was shown to be near 100%. Most gas phase ozone,
NO2 and ammonia is also removed by the denuder (Ding 1999b; Eatough
1999; Obeidi 1999). The agreement for fine particulate mass, carbonaceous
material, sulfate and nitrate between collocated PC-BOSS samplers was
±8% and agreement for these species between the PC-BOSS and comparable
diffusion denuder samplers was ±5-10% (Ding 1999b, Pang 1999a,b). For
all studies, the results obtained with the PC-BOSS sampler indicated
that from 10 to 50% of the fine particulate mass was not measured with
the PM2.5 FRM sampler due to the loss of semi-volatile organic material
and ammonium nitrate during sampling. The majority of the loss at Riverside
and Provo was due to semi-volatile organic material, but at Bakersfield
the major species lost was ammonium nitrate.
During
the intensive periods samples five samples will be collected and analyzed
daily with a PC-BOSS sampler by the BYU team to establish the diurnal
pattern of fine particulate chemical species, including semi-volatile
nitrate and organic material. Approximately half of the samples collected
during the three two-week intensive field sampling programs will be
selected for analysis. PM2.5 mass will be determined from the material
collected on the Teflon filter in the minor channel of the PC-BOSS (Figure
A3). This mass will be corrected for the loss of ammonium nitrate and
semi-volatile organic material (determined from analysis of the Nylon
and charcoal impregnated filter, CIF, in the minor flow channel of the
PC-BOSS). Constructed mass will also be determined from the various
analyzed species. The concentration of fine particulate sulfate and
of nitrate retained by a particle collection filter will be determined
by ion chromatographic analysis of ultrasonic aqueous extracts of each
quartz and Teflon filter (a total of three filters/sample) of every
PC-BOSS collected sample. The concentration of nitrate lost from particles
during sample collection will be obtained by ion chromatographic analysis
of IC eluent extracts of both the Nylon and CIF filters of the PC-BOSS.
The concentrations of ammonium ion in extracts of the minor flow Teflon
filter for each PC-BOSS sample will be determined spectrophotometrically.
The pH of these extracts will be determined by pH measurement and the
acidity obtained as previously described (Koutrakis et al., 1988). The
concentrations of organic carbonaceous material and soot retained on
each PC-BOSS quartz filter (Figure A3), will be determined by temperature
programmed volatilization analysis (TPV) from ambient to 800 EC in a
N2/O2 atmosphere (Eatough 1993, Ellis 1982). The concentrations of semi-volatile
organic material lost from the particles during sampling will be obtained
from TPV analysis of the material collected by each CIF of the PC-BOSS
in a N2 atmosphere from ambient to 400 EC (Eatough 1993, Tang 1994).
2.7
Continuous Nitrate, Sulfate, and Carbon Measurements
For
particle nitrate the ADI approach has been compared to denuder-filter
methods at three different California locations, namely Riverside, Mira
Loma and Bakersfield, and has yielded regression slopes of 0.96, 1.00
and 1.07 with correlation coefficients of 0.96. Example data from Bakersfield,
California is shown in Figure A4 (Dutcher et al., 1999). Laboratory
tests show collection efficiencies of greater than 95% for particles
between 0.1-1 µm. Interference testing shows less than 1% response to
nitric acid or ammonium ion.
Data
from the combined sulfate, carbon nitrate system collected in December
1998 in Marshall, Colorado are shown in Figure A5 (Hering et al, 1999).
The data show that the time
|
|
| Figure A4.
Comparison of nitrate concentrations measured by the automated nitrate
monitor with measured by a filter sampler. |

Figure
A5. Time series of the automated sulfate, nitrate, carbon measured at
Marshall, Colorado, with comparison of the mass sum of species (left
hand axis) to optical particle counter volume (mm3/cc, right hand axis).
Sum of species is calculated as 1.29*nitrate + 1.375*sulfate +1.4*carbon.
variation
in the sum of the nitrate, sulfate, and carbon species measured by these
systems tracked the total particle concentration, as indicated by physical
size distribution measurements. Additionally, the mass sum is comparable
to the measured aerosol volume, with a mass to volume ratio just over
1 g cm-3. Note that the relative proportions of carbon, sulfate, and
nitrate vary. Such data are extremely valuable to the proposed atmospheric
aerosol modeling for the Pittsburgh Supersite. The combined modeling
and measurement effort will provide valuable insights into the sources
and formation processes for these particle constituents which comprise
the majority of the fine particle mass.
2.8
Bioaerosols
The
Hernandez group using direct microscopy and molecular biological analyses
will develop and apply rapid, quantitative, aerosol assays to characterize
the identity, distribution and activity of microbiological components
present in outdoor aerosols. This aerosol microbiology study will focus
on bacteria, fungi and their spores. The definition of "microbiological"
here does not dictate the necessity to be isolated or quantified by
culture. In response to the limitations of conventional microbiological
analysis, the Hernandez group will refine accepted (and novel) epi-fluorescent
microscopy practices with genetic-based, molecular biology assays into
rapid, non-culture based techniques that can directly measure and classify
microbiological bioaerosol material in outdoor air. Our sampling approach
for characterizing outdoor bioaerosols is to use two different types
of common samplers in conjunction with each other: cyclone wet samplers
and high volume PM10&2.5 filter samplers. The wet samplers will be used
for culturing, microscopic and genetic analyses, while the total PM
samplers will be used for biomass determinations (vs. total mass). Sensitive
biological stains and image analyses will be used to directly characterize
microbiological aerosols collected by both types of samplers. Image
analysis will be used to accurately measure the size and morphology
of stained biological particle recovered from outdoor air. Using widely
accepted molecular methods, genetic probes, polymerase chain reaction
(PCR), and denaturing gradient gel electrophoresis (DGGE) will be used
to classify airborne microorganisms on a phylogenic, rather than a culturable
basis.
2.9
Meteorology
Atmospheric
trajectory analysis has long been used to identify the location of upwind
regions that contribute material to samples collected hundreds and even
thousands of kilometers downwind. Information contained in modeled trajectories
includes the upwind transport pathway corresponding to air arriving
at the sampling site and the temperature at each point in the trajectory.
This information identifies the transit time from upwind sources, thus
aiding in the determination of en route chemical transformation and
possible interaction with clouds, among other things. The trajectory
model developed by Harris and Kahl (1994) will be applied by the Kahl
group to calculate trajectories twice daily for a 1.5 year segment of
the project. This model calculates 10-day transport along isentropic
surfaces, thus accounting for the adiabatic vertical motions that characterize
the basic features of atmospheric transport. The model switches from
isentropic to mixed-layer mode when trajectories come close to the earth's
surface. In this mode, air parcels are advected using winds averaged
throughout the lowest 500 m of the atmosphere. In the isentropic mode
the parcels are advected using winds found along a surface of constant
potential temperature. The model also features a dynamic "theta-chooser"
algorithm, which calculates the potential temperature surface that corresponds
to the desired altitude of trajectory arrival at the receptor.
2.10
Gas-phase measurements
Individual
VOC concentrations will be measured chromatographically using the GC-FID
and GC/MS techniques. Samples will be collected into 32 L Summa canisters
that will be pressurized during sampling to approximately the equivalent
of 96 L of air at one atmosphere. The GC-FID method will use an HP gas
chromatograph and integrator system with a J&W Scientific DB-1 column.
The sample will be loaded through a gas-sampling valve onto a Supelco
Carbotrap 300 sorbent trap, at 200 to 300 mL min-1. Residual water will
be removed by passing He through the trap at 100 mL min-1 for 7 min.
The trapped sample is then desorbed at 340oC for 10 min through a second
gas-sampling valve into a loop filled with glass beads. The organics
are vaporized by heating the loop at 100oC and cryo-focused onto the
head of the column held at -60oC. The column-oven temperature is held
for 5 min at -60oC and then increases at 4oC min-1 to 200oC. Using a
similar technique Lewis et al. (1999) reported a species quantification
limit of about 0.1 µg m-3 allowing them to quantify the concentrations
of roughly 130 compounds in Atlanta air.
The
GC-MS analysis will be similar to the GC-FID. Ambient samples will be
loaded onto a multibed Tenax TA-Ambersorb-Charcoal sorbent trap in a
Dynatherm Analytical Instruments model 890. The sample will be desorbed
at 300oC for 6 minutes to a freeze-out loop immersed in liquid nitrogen.
The rest of the analysis will follow the GC-FID protocol.
The
Colorado State University (CSU) team will make gas phase measurements
of hydrogen peroxide and soluble organic peroxides using a continuous
monitor based on the method of Lazrus et al. (1986). This dual-channel
instrument, constructed at CSU with the assistance of Drs. Greg Kok
and Teresa Campos of NCAR, features a detection limit of ~ 0.1 ppbv.
Air is drawn into the instrument through a Teflon sample tube. The first
instrument channel scrubs soluble hydroperoxides (including hydrogen
peroxide and soluble organic peroxides) from the air into solution where
they are reacted to produce a fluorescent dimer, the concentration of
which is measured by an on-line fluorimeter. The second instrument channel
is identical to the first except that catalase is added to the sample
stream to destroy hydrogen peroxide, permitting measurement of residual,
organic peroxide concentrations. The hydrogen peroxide concentration
is given by the difference between the two channel concentrations. The
peroxide analyzer is calibrated by automated injection of aqueous standards.
Routine checks are also made to verify efficient transmission of gaseous
peroxides through the Teflon sample tube and efficient destruction of
hydrogen peroxide by catalase in the second channel. Data will be recorded
at 10 minute intervals and averaged to longer periods as needed for
use in the cloud/fog study and for tests of the influence of hydrogen
peroxide and organic peroxides on health. Expertise and instrumentation
for the peroxide measurements will be provided by CSU with on-site assistance
from CMU personnel.
2.11
Polar Organics
Polar
organic compounds will be detected by an aerosol MALDI experiment. Matrix-assisted
laser desorption ionization (MALDI) is widely used to obtain mass spectra
of highly polar compounds. The analysis is performed by mixing the analyte
with a matrix compound that strongly absorbs the laser radiation. The
matrix-analyte mixture is then irradiated with a pulsed laser beam to
eject intact analyte ions without fragmentation. The U. of Delaware
group has performed MALDI of bulk and single-particle samples for several
years (e.g., Mansoori et al., 1996).
On-line
single particle MALDI will be performed as follows. Ambient aerosol
will be mixed with a small flow of the hot vapor matrix quickly cooling
it to ambient temperature and reducing the vapor pressure. The matrix
vapor will condense on existing aerosol particles producing a matrix/analyte
mixture like that used in a conventional MALDI experiment. This principle
is similar to that of a "mixing-type" condensation nucleation counter
(Okuyama et al, 1984) except that a much smaller amount of vapor will
be condensed on the particles. The increase in particle size due to
matrix condensation is controlled by the relative flow of the matrix
vapor and the aerosol. A large body of literature has established the
operating parameters (e.g., Willeke and Baron, 1993).
The
U. of Delaware team has built a prototype apparatus to condense a very
thin film of benzoic acid matrix on ultrafine particles. The goal of
this experiment was not to perform MALDI of aerosol organics but rather
to study the effect of a thin, absorbing coating on the yield of ions
from non-absorbing particles. With this apparatus, it was found that
a benzoic acid film corresponding less than a 5% increase in the particle
size (the smallest size increase that could be measured with our experimental
setup) was sufficient to increase the number of particles analyzed by
as much as a factor of two. This increase in the "hit" rate arises from
particles that pass through the edge of the laser beam. Because particles
near the edge of the laser beam experience a lower laser irradiance
than particles in the center of the beam, they may yield too few ions
to be detected. Adding a small amount of benzoic acid increases the
amount of laser radiation that is absorbed by the particle and increases
the yield of ions from the core material. In this way, particles near
the edge of the laser beam are detected and the hit rate increases.
This experiment underscores the dramatic effect that even a small amount
of an absorbing material can have on the laser ablation process.
Based
upon conventional MALDI experiments, we expect that no more than a 1:1
mass ratio of matrix to analyte is needed to enhance formation of organic
molecular ions. Since particle mass is proportional to the cube of the
diameter, a doubling of the particle mass by condensation of matrix
vapor will result in a 26% increase in particle diameter. This size
change is small relative to the density uncertainty and corresponding
size uncertainty for aerodynamically sized particles and therefore should
not significantly alter conclusions drawn from correlated size-composition
measurements of the transformed particles.
2.12
Semi-continuous metals
The
semi-continuous metals sampler of the Ondov group consists of a high-frequency
aerosol sampler (HFAS) and state-of-the-art true simultaneous multi-element
Graphite Furnace Atomic absorption spectrometer. The HFAS consist of
a state-of-the-art dynamic aerosol concentrator. PM 2.5 is sampled at
200 L min-1 and delivered to GFAA system or to a sample fraction collector
for on or off-line analyses, respectively. In <10 (for urban air) to
20 (rural air) minutes the U. of Maryland (UMCP) group typically collects
enough slurry to permit 4 suites of 4 or 5 elements to be determined,
each in triplicate. At the 200 L min-1 sampling rate, analyte masses
delivered to the GFAA at the smallest concentrations observed by Ondov
in rural Maryland air exceed instrumental sensitivities (rivaling far
more expensive ICP-MS but with far less sample volume requirement) by
factors of 2 to >400 for Al, As, Cd, Cu, Fe, Mn, Ni, Pb, Se, V, and
Zn. In addition to high temporal resolution, it is of paramount importance
that analytical concentration measurements be accurate. Tests with NIST
Standard Reference Material 1648 ("Urban Particulate Material") confirm
these results for Cd, Pb, Zn, Se, As, Cr, Mn, Cu, and Ni. Tests for
the remaining elements are in progress. Additional field tests of the
instrument are being made in College Park, MD, and at the Atlanta Super
Site in August, 1999.
The
Ondov group will construct and deploy two HFASs, i.e., one at the main
site and one at the satellite site. These will be deployed for 2 weeks
during each of 3 two-week intensives and will be used to collect samples
continuously at a sub-hourly interval to be selected during a pilot
field test prior to the first scheduled intensive. To save costs and
permit a maximum number of analyses to be performed, samples will be
collected in fraction collectors (each holding up to 288 samples) and
returned to UMCP for analysis on a (truly) simultaneous Graphite-Furnace
Atomic Absorption Spectrometer for 18 elements in the following four
groups:
Group
1: As, Cu, Mn, Ni, Cr
Group 2: Cd,
Se, Ag, Pb
Group 3: Al, Fe, Zn, Ca, and Bi (Bi is an internal standard),
Group 4: V, Ti, Be, Ba
The
elements selected include criteria pollutants (Pb and Be), hazardous
air pollutants (known as air toxins; i.e., Cd, Cr, Cu, Ni, As, and Se),
first series transition metals either known (V and Zn) or suspected
to elicit respiratory inflammation (Ti, V, Cr, Mn, Fe, Ni, Cu, Zn; i.e.,
all but Sc and Co); essential nutrients (Fe, Zn, Se, Cr), and the aquatic
toxin, Al. In addition, most are primary source marker species (i.e.,
As and Fe, steel; Zn, Cd, Cr, Pb, Ag, incinerators; Ni and V, residual
fuel oil combustion; Se, coal combustion; Ti, paint manufacture and
applications (fine particles) or crustal dust (coarse particles); Ca
limestone/construction material; and Cu, Cd, Zn, respective smelter
emissions and/or metals processing.
Until
more experience is gained, 3 replicate analyses will be made for each
suite of elements. Analysis of each suite about 5 minutes per replicate,
so that the time to complete 3 replicates is 15 minutes per suite. The
initial target collection rate will be 6 samples per hour, pending ambient
concentrations. As the collection rate will far exceed the analytical
rate the UMCP group will analyze selected series of samples based on
observations made with continuous aerosol mass and particle concentrations,
SO2 concentration, and wind direction. At 200 samples from each intensive
will be analyzed. Many more samples can be analyzed if fewer suites
are selected, however, the best strategy may be to remove replications
as 10 minute sampling times will likely provide for replication, in
which case >600 samples would be analyzed per intensive. Note that each
fraction collector can accommodate 288 samples, i.e., about 2 days worth.
2.13
Three-Dimensional Modeling Study
In
collaboration with Ted Russell in Georgia Tech., the CMU groups have
created a comprehensive three-dimensional model for the study of PMx
in the region shown in Figure A6. The model includes a state-of-the-art
description of PM processes and describes the complete aerosol size/composition
distribution using user-selected chemical and size resolution. The model
is also coupled to a sensitivity analysis module so it can calculate
directly the sensitivities of PM concentrations to small changes in
source strength. The response to larger changes can be calculated with
additional simulations.

Figure
A6. Modeling region used in current simulations of the size/composition
distribution of PM in the central and eastern US by the CMU group. The
model grid can be adjusted for different applications. High resolution
(5x5 km) not shown here will be used in the area surrounding the Pittsburgh
Supersite for additional simulations.